Analyze Waveform Data
For example, to use
Edit Expression
to take the integral of a square wave:
1.
Push the lower-bezel
Clear
button.
2.
Turn multipurpose knob
a
to select
Intg(
.
3.
Push
Enter Selection
.
4.
Turn multipurpose knob
a
to select
channel
1
.
5.
Push
Enter Selection
.
6.
Turn multipurpose knob
a
to select
)
.
7.
Push
OK Accept
.
Using Reference Waveforms
Create a reference waveform to store a waveform. For example, you might do this to set up a standard against which to compare
other waveforms. To use the reference waveforms:
NOTE.
10 M reference waveforms are volatile and not saved when the oscilloscope power is turned off. To keep these waveforms,
save them to external storage.