Display Waveform Data
9.
Select
Offset
to enable multipurpose knob
a
to make vertical offset adjustments.
On the side-bezel menu, choose
Set to 0 V
to
set the vertical offset to 0 V.
For more information on offset, see
Quick
Tips
10.
Select
Probe Setup
to de
fi
ne probe
parameters.
On the resulting side-bezel menu:
Select
Voltage
or
Current
to set the
probe type for probes that do not have the
TekProbe II or TekVPI interface.
Use multipurpose knob
a
to set the
attenuation to match the probe.
11.
Select
Deskew
to set the time skew correction
for the channel. Turn multipurpose knob
a
to
adjust the time skew (deskew) correction for
the probe attached to the selected channel.
This shifts acquisition and display of the
waveform left or right, relative to the trigger
time. Use this to compensate for differences in
cable lengths or probe types.
Tektronix 4000 Series Oscilloscopes User Manual