Installation
Acquisition System: 1 M
Ω
The maximum input voltage at the BNC, between center conductor and shield is 400 V
pk
(DF
≤
39.2%), 250 V
RMS
to 130 kHz
derated to 2.6 V
RMS
at 500 MHz.
The maximum transient withstand voltage is ± 800 V
pk
.
For steady-state sinusoidal waveforms, derate at 20 dB/decade above 200 kHz to 13 V
pk
at 3 MHz and above.
Acquisition System: 50
Ω
The maximum input voltage at the BNC, between center conductor and shield is 5 V
RMS
, with peaks
≤
±20 V (DF
≤
6.25%)
Acquisition System: Digital Inputs
The maximum input voltage at the input for the logic probe is ±15 V peak.
Aux In: 1 M
Ω
The maximum input voltage at the BNC, between center conductor and shield is 400 V
pk
(DF
≤
39.2%), 250 V
RMS
to 2 MHz
derated to 5 V
RMS
at 500 MHz.
The maximum transient withstand voltage is ±800 V
pk
.
For steady-state sinosoidal waveforms, derate at 20 dB/decade above 200 kHz to 13 V
pk
at 3 MHz and above.
CAUTION.
To ensure proper cooling, keep the sides and rear of the instrument clear of obstructions.