Trigger Setup
10.
If you are setting up an RS-232 trigger and
have made a
Trigger On
selection of
Tx Data
or
Rx Data
, push the lower-bezel
Data
button.
Push the
Number of Bytes
button and enter
the number of bytes with multipurpose knob
a
.
Push the side-bezel
Data
button, and enter
the parameters of interest with multipurpose
knobs
a
and
b
.
I
2
C, SPI, and CAN Bus Trigger Data Matching
Rolling window byte matching for I
2
C and SPI.
To use a rolling window to trigger on data, you de
fi
ne the number of bytes to
match. Then the oscilloscope uses a rolling window to
fi
nd any match within a packet, with the window rolling one byte at a time.
For example, if the number of bytes is one, the oscilloscope will attempt to match the
fi
rst byte, second byte, third, and so
on within the packet.
If the number of bytes is two, the oscilloscope will try to match any two consecutive bytes, such as one and two, two and three,
three and four, and so on. If the oscilloscope
fi
nds a match, it will trigger.
Speci
fi
c byte matching (non-rolling window matching for a speci
fi
c position in the packet) for I
2
C, SPI, and
CAN.
You can trigger on a speci
fi
c byte for I
2
C, SPI, and CAN in several ways:
For I
2
C and SPI, enter the number of bytes to match the number of bytes in the signal. Then use don’t cares (X) to mask the
bytes that you are not interested in.
Tektronix 4000 Series Oscilloscopes User Manual
139
Summary of Contents for 4000 Series
Page 4: ......
Page 18: ...Preface xiv Tektronix 4000 Series Oscilloscopes User Manual ...
Page 53: ...Installation 2 Power off your oscilloscope Tektronix 4000 Series Oscilloscopes User Manual 35 ...
Page 68: ...Installation 50 Tektronix 4000 Series Oscilloscopes User Manual ...
Page 312: ...Application Examples 294 Tektronix 4000 Series Oscilloscopes User Manual ...