Trigger Setup
Trigger Type
Trigger Conditions
Setup and Hold
Trigger when a logic data input changes state inside of the
setup or hold time relative to a clock edge.
Setup is the amount of time that data should be stable and
not change before a clock edge occurs. Hold is the time
that data should be stable and not change after a clock
edge occurs.
MSO4000 series oscilloscopes are capable of multiple
channel Setup and Hold triggering, and can monitor the
state of an entire bus for setup and hold violations. You
can use up to 20 channels for a Setup and Hold trigger (4
analog and 16 digital) with MSO4000 series oscilloscopes.
Use the Clock control side-bezel button to select the clock
channel. Use the Select control and the Function control
(MSO4000 series only) side-bezel buttons to select one
or more channels you want to monitor for setup and hold
violations.
NOTE.
Optimum Setup and Hold trigger performance
is achieved by using only analog channels or only digital
channels.
Tektronix 4000 Series Oscilloscopes User Manual
129