Appendix: Warranted Speci
fi
cations, Safety Certi
fi
cations, and Electromagnetic Compatibility
LOW TRUE; LOW to HIGH transition indicates that the trigger occurred. The logic levels are given
in the following table:
Characteristic
Limits
Vout (HI)
≥
2.5 V open circuit;
≥
1.0 V into a 50
Ω
load
to ground
Auxiliary output (AUX
OUT)
Vout (LO)
≤
0.7 V into a load of
≤
4 mA;
≤
0.25 V into a
50
Ω
load to ground
Digital Channel
Threshold Accuracy,
MSO4000 series
± [100 mV + 3% of the threshold setting after calibration]
Safety Certi
fi
cation
Listed UL61010–1: 2004, CAN/CSA-C22.2 No. 61010.1: 2004;
Complies with EN61010–1: 2001,
Complies with the Low-Voltage Directive 73/23/ECC for Product Safety
Electromagnetic Compatibility (EMC)
Tektronix 4000 Series Oscilloscopes User Manual