ENDOCAPSULE 10 SYSTEM

This quick reference guide is only a summary of the ENDOCAPSULE instructions for use. For detailed operating instructions, 
be sure to follow the ENDOCAPSULE instruction manual that was included with your device when purchased. 

Preparation and settings for capsule endoscopy

 1. Place recorder in the cradle.

3.  Insert antenna sheet into the 

cover. Orange colored lines 
printed on the surface should 
face front. The cable always 
stays on the patient’s left side.

7.  Insert the capsule activator into 

the capsule container. The LED 
should start blinking.

11.  Unlock the recorder. Press “MENU” for more 

than 2 sec. Open “Real Time View” to confirm 
that the examination is complete.

12.  Place recorder in the cradle.

13.  A message will be displayed on the screen. 

Click “OK” to start downloading image data. 
When completed, a check mark will be 
displayed on the main screen.

4.  Fasten the antenna belt 

around the patient’s body. The 
alignment hole should fit the 
umbilical area.

8.  Press “MENU” and select “Real 

Time View”. Confirm that images 
are received properly.

 5.  Attach the recorder holder on 

the left side of the patient’s 
body. Connect the recorder 
and the antenna unit. Slide the 
connection part until it clicks.

6.  Press “POWER” for at least 

1 sec. Patient ID and name 
are displayed on the upper left 
screen.

 2. Log in to workstation. Register 
patient information and initialize 
recorder.

 

Refer to the back page 

3

7

11

12

13

3

3

6

Before the examination

After the examination

1

2

Note: Check that the battery is 
fully charged (green light is lit).

IMPORTANT NOTICE 
TO PATIENTS 

When removing the capsule 
from the container, be careful 
NOT to touch the top cover of 
the capsule. Otherwise, clear 
images may not be captured.

8

4

 9.  Press “MENU” and select 

“Real Time View”. Confirm that 
images from the body are being  
received properly. If necessary, 
images can be played back or 
captured.

10.  Lock the button operation. Go to “Patient 

Message” and press “MENU” for at least 
2 sec. When patient guidance is activated, 
the patient does not necessarily need to 
touch the recorder. Beep or vibration will 
stop after 30 sec.

9

10

After the capsule ingestion

5