Vertical Controls
3- 84
TDS3000B Series User Manual
50
Ω
Protection.
If you select the 50
Ω
termination resistance, the
maximum vertical scale factor is limited to 1 V/div. If you apply
excessive input voltage, the oscilloscope automatically switches to
1 M
Ω
termination resistance to protect the internal 50
Ω
termination.
Math Button
Push the MATH button to define the math waveform using the math
menu. Also push the MATH button to display or select the math
waveform. The bottom menu may show other menu items depending
on which application module you have installed.
Bottom
Side
Description
Dual Wfm Math
Set 1st Source to
Selects the first source waveform.
Set Operator to
Selects the math operator: +, --,
×
,
or
÷
Set 2nd Source to Selects the second source waveform.
Key Points
Dual-Waveform Math.
For the dual-waveform math operations, the two
source waveforms interact with the math operators in the sequence
shown below.
Operation
Expression of math waveform
+
Source 1 + Source 2
--
Source 1 -- Source 2
×
Source 1
×
Source 2
÷
Source 1
÷
Source 2
Summary of Contents for 071-0957-03
Page 4: ......
Page 15: ...Getting Started ...
Page 16: ......
Page 51: ...Application Examples ...
Page 52: ......
Page 80: ...Application Examples 2 28 TDS3000B Series User Manual ...
Page 81: ...Reference ...
Page 82: ......
Page 176: ...e Scope 3 94 TDS3000B Series User Manual ...
Page 177: ...Appendices ...
Page 178: ......
Page 216: ...Appendix E Performance Verification E 2 TDS3000B Series User Manual ...
Page 218: ...Appendix F General Care and Cleaning F 2 TDS3000B Series User Manual ...
Page 234: ...Appendix G Ethernet Setup G 16 TDS3000B Series User Manual ...
Page 235: ...Glossary ...
Page 236: ......
Page 247: ...Index ...
Page 248: ......