Appendix G: Ethernet Setup
G- 2
TDS3000B Series User Manual
However, if you intend to remotely control or access the oscilloscope,
have the system administrator assign a
static
IP address so that the
oscilloscope IP address does not change. A static IP address makes it
easier for remote devices to access a specific oscilloscope.
Entering the Ethernet Network Settings
The procedure for entering the oscilloscope Ethernet network
parameters depends on your network configuration. If your network
supports DHCP or BOOTP, then use the following procedure. If your
network does not support DHCP or BOOTP, then use the procedure
on page G--3.
Networks That Support DHCP and BOOTP
If your network supports DHCP/BOOTP, do these steps:
1.
Push the
UTILITY
front panel button.
2.
Push the
System
menu button to select
I/O
.
3.
Push the
Ethernet Network Settings
screen button.
4.
Push the
DHCP/BOOTP
side button to select
On
. The screen
displays the clock icon while it is talking with the network to
obtain an IP address for the oscilloscope. This step should only
take a few moments, but the actual time will vary depending on
your network. The clock icon disappears when the task is
finished.
Summary of Contents for 071-0957-03
Page 4: ......
Page 15: ...Getting Started ...
Page 16: ......
Page 51: ...Application Examples ...
Page 52: ......
Page 80: ...Application Examples 2 28 TDS3000B Series User Manual ...
Page 81: ...Reference ...
Page 82: ......
Page 176: ...e Scope 3 94 TDS3000B Series User Manual ...
Page 177: ...Appendices ...
Page 178: ......
Page 216: ...Appendix E Performance Verification E 2 TDS3000B Series User Manual ...
Page 218: ...Appendix F General Care and Cleaning F 2 TDS3000B Series User Manual ...
Page 234: ...Appendix G Ethernet Setup G 16 TDS3000B Series User Manual ...
Page 235: ...Glossary ...
Page 236: ......
Page 247: ...Index ...
Page 248: ......