Application Examples
2- 6
TDS3000B Series User Manual
Customizing Your Measurements
In this example you want to verify that the incoming signal to a
piece of digital equipment meets its specifications. Specifically, the
transition time from a low logic level (0.8 V) to a high logic level
(2.0 V) must be 10 ns or less.
2.0 V
≤
10 ns
0.8 V
To select the rise time measurement, do these steps:
1.
Push the
MEASURE
button to see the measurement menu.
2.
Push the
CH 1
button and then the
Select Measurement for Ch1
screen button.
3.
Select the
Rise Time
measurement.
Rise time is typically measured between the 10% and 90% amplitude
levels of a signal; these are the default reference levels the
oscilloscope uses for rise time measurements. However, in this
example you need to measure the time the signal takes to pass
between the 0.8 V and 2.0 V levels.
You can customize the rise time measurement to measure the signal
transition time between any two reference levels. You can set each of
those reference levels to a specific percent of the signal amplitude or
to a specific level in vertical units (such as volts or amperes).
Summary of Contents for 071-0957-03
Page 4: ......
Page 15: ...Getting Started ...
Page 16: ......
Page 51: ...Application Examples ...
Page 52: ......
Page 80: ...Application Examples 2 28 TDS3000B Series User Manual ...
Page 81: ...Reference ...
Page 82: ......
Page 176: ...e Scope 3 94 TDS3000B Series User Manual ...
Page 177: ...Appendices ...
Page 178: ......
Page 216: ...Appendix E Performance Verification E 2 TDS3000B Series User Manual ...
Page 218: ...Appendix F General Care and Cleaning F 2 TDS3000B Series User Manual ...
Page 234: ...Appendix G Ethernet Setup G 16 TDS3000B Series User Manual ...
Page 235: ...Glossary ...
Page 236: ......
Page 247: ...Index ...
Page 248: ......