Acquisition Controls
3- 6
TDS3000B Series User Manual
Acquire Menu
Push the Acquire MENU button to show the acquire menu.
Bottom
Side
Description
Mode
Sample
Use for normal acquisition.
Peak Detect
Detects glitches and reduces the
possibility of aliasing.
Envelope N
Captures variations of a signal over a
period of time. (Adjust N with the
general purpose knob.)
Average N
Reduces random or uncorrelated noise
in the signal display. (Adjust N with the
general purpose knob.)
Horizontal
Resolution
Fast Trigger
(500 points)
Acquires 500 point waveforms at a fast
repetition rate.
Normal
(10k points)
Acquires 10,000 point waveforms with
more horizontal detail.
Reset Horizontal
Delay
Set to 0 s
Resets the horizontal delay time to zero.
Autoset
Normal Autoset
Executes the autoset function. (Optional
application modules may add choices to
execute specialized autoset functions.)
Undo Autoset
Reverts to the settings before the last
autoset.
Summary of Contents for 071-0957-03
Page 4: ......
Page 15: ...Getting Started ...
Page 16: ......
Page 51: ...Application Examples ...
Page 52: ......
Page 80: ...Application Examples 2 28 TDS3000B Series User Manual ...
Page 81: ...Reference ...
Page 82: ......
Page 176: ...e Scope 3 94 TDS3000B Series User Manual ...
Page 177: ...Appendices ...
Page 178: ......
Page 216: ...Appendix E Performance Verification E 2 TDS3000B Series User Manual ...
Page 218: ...Appendix F General Care and Cleaning F 2 TDS3000B Series User Manual ...
Page 234: ...Appendix G Ethernet Setup G 16 TDS3000B Series User Manual ...
Page 235: ...Glossary ...
Page 236: ......
Page 247: ...Index ...
Page 248: ......