Acquisition Controls
3- 12
TDS3000B Series User Manual
Acquisition Resolution.
You can choose either Normal or Fast Trigger
acquisition resolution. This setting determines the record length that
is acquired and affects the other factors shown in the table below.
Factor
Normal
Fast trigger
Record length
10,000 points
500 points
Maximum acquisition rate:
300 MHz -- 600 MHz models
100 MHz -- 200 MHz models
700 waveforms/s
400 waveforms/s
3,600 waveforms/s
2,600 waveforms/s
Maximum horizontal zoom factor
200X
10X
Choose Normal or Fast Trigger acquisition resolution based on the
characteristics of the signal you want to acquire.
Signal characteristic
Better choice
Large amount of horizontal detail
Normal
Shape is stable or changes relatively slowly
Normal
Single shot
Normal
High trigger repetition rate
Fast Trigger
Shape changes rapidly
Fast Trigger
Contains modulation
Fast Trigger
Summary of Contents for 071-0957-03
Page 4: ......
Page 15: ...Getting Started ...
Page 16: ......
Page 51: ...Application Examples ...
Page 52: ......
Page 80: ...Application Examples 2 28 TDS3000B Series User Manual ...
Page 81: ...Reference ...
Page 82: ......
Page 176: ...e Scope 3 94 TDS3000B Series User Manual ...
Page 177: ...Appendices ...
Page 178: ......
Page 216: ...Appendix E Performance Verification E 2 TDS3000B Series User Manual ...
Page 218: ...Appendix F General Care and Cleaning F 2 TDS3000B Series User Manual ...
Page 234: ...Appendix G Ethernet Setup G 16 TDS3000B Series User Manual ...
Page 235: ...Glossary ...
Page 236: ......
Page 247: ...Index ...
Page 248: ......