TDS3000B Series User Manual
A- 1
Appendix A: Specifications
This appendix contains specifications for the TDS3000B Series
oscilloscopes. All specifications are guaranteed unless noted as
“typical.” Typical specifications are provided for your convenience
but are not guaranteed. Specifications that are marked with the
n
symbol are checked in
Appendix E: Performance Verification
.
All specifications apply to all TDS3000B Series models unless noted
otherwise. To meet specifications, two conditions must first be met:
H
The oscilloscope must have been operating continuously for
twenty minutes within the operating temperature range specified.
H
You must perform the Compensate Signal Path operation
described on page 1--4. If the operating temperature changes by
more than 10
°
C, you must perform the Compensate Signal Path
operation again.
Specifications
Acquisition
Acquisition modes
Sample (Normal), Peak detect, Envelope, and Average
Single Sequence
Acquisition mode
Acquisition stops after
g
q
Sample, Peak Detect
One acquisition, all channels
simultaneously
Average, Envelope
N acquisitions, all channels
simultaneously, N is settable
from 2 to 256 (or
∞
for Enve-
lope)
Summary of Contents for 071-0957-03
Page 4: ......
Page 15: ...Getting Started ...
Page 16: ......
Page 51: ...Application Examples ...
Page 52: ......
Page 80: ...Application Examples 2 28 TDS3000B Series User Manual ...
Page 81: ...Reference ...
Page 82: ......
Page 176: ...e Scope 3 94 TDS3000B Series User Manual ...
Page 177: ...Appendices ...
Page 178: ......
Page 216: ...Appendix E Performance Verification E 2 TDS3000B Series User Manual ...
Page 218: ...Appendix F General Care and Cleaning F 2 TDS3000B Series User Manual ...
Page 234: ...Appendix G Ethernet Setup G 16 TDS3000B Series User Manual ...
Page 235: ...Glossary ...
Page 236: ......
Page 247: ...Index ...
Page 248: ......