Calibration System
Use the System Cal branch to access these functions.
Bottom
Side
Description
System Cal
Signal Path
Compensates the signal paths to
obtain best measurement accuracy.
Factory Cal
Used to calibrate the oscilloscope.
This is a service function only.
Cal Due Control
Notify After Hours
of Operation
Sets the number of hours of operation
before notifying you that a calibration
is due.
Notify After Years
of Elapsed Time
Sets the number of years before
notifying you that a calibration is due.
Key Points
Signal Path Compensation.
For maximum accuracy at any time, run the
signal path compensation routine just before taking critical
measurements. To meet accuracy specifications, run the routine if
the ambient temperature changes by 10
°
C or more.
Before running the routine, disconnect any probes or cables from the
channel inputs. Then, push the Signal Path and OK Compensate
Signal Path screen buttons to confirm that you are ready to proceed.
The routine takes several minutes to complete.
UTILITY
Summary of Contents for 071-0957-03
Page 4: ......
Page 15: ...Getting Started ...
Page 16: ......
Page 51: ...Application Examples ...
Page 52: ......
Page 80: ...Application Examples 2 28 TDS3000B Series User Manual ...
Page 81: ...Reference ...
Page 82: ......
Page 176: ...e Scope 3 94 TDS3000B Series User Manual ...
Page 177: ...Appendices ...
Page 178: ......
Page 216: ...Appendix E Performance Verification E 2 TDS3000B Series User Manual ...
Page 218: ...Appendix F General Care and Cleaning F 2 TDS3000B Series User Manual ...
Page 234: ...Appendix G Ethernet Setup G 16 TDS3000B Series User Manual ...
Page 235: ...Glossary ...
Page 236: ......
Page 247: ...Index ...
Page 248: ......