Application Examples
TDS3000B Series User Manual
2- 7
Setting Reference Levels.
To set the reference levels to specific
voltages, do these steps:
1.
Push the
Reference Levels
screen button.
2.
Push the
Set Levels in
screen button to select
units
.
3.
Push the
High Ref
screen button.
4.
Use the general purpose knob to select
2.0 V
.
5.
Push the
Low Ref
screen button.
6.
Use the general purpose knob to select
800 mV
.
The measurement verifies that the transition time (3.842 ns) meets
the specification (
≤
10 ns).
Measuring Specific Events.
Next you want to see the pulses in the
incoming digital signal, but the pulse widths vary so it is hard to
establish a stable trigger. To look at a snapshot of the digital signal,
do this step:
1.
Push the
SINGLE SEQ
button to capture a single acquisition.