![Tektronix 071-0957-03 User Manual Download Page 158](http://html.mh-extra.com/html/tektronix/071-0957-03/071-0957-03_user-manual_1078806158.webp)
Utility
3- 76
TDS3000B Series User Manual
RS-232 Troubleshooting.
If you are having difficulties with RS-232
communication, try the following remedies:
H
Verify that you are using the correct RS-232 cable and adapters.
Most computers require a null-modem connection to the
oscilloscope. Most printers require a straight-through connection
to the oscilloscope.
H
Verify that the RS-232 cable is connected to the correct port on
your computer or hard copy device.
H
Reset the RS-232 parameters to defaults and then set the baud
rate to match the computer or hard copy device. The default
settings (except baud rate) are standard on most computers and
hard copy devices.
H
Enable the debug window to see the RS-232 status, errors, data
transmitted, and data received.
GPIB Guidelines.
Follow these guidelines when you connect your
oscilloscope to a GPIB network:
H
Turn off the oscilloscope and all external devices before
connecting the oscilloscope to the GPIB network.
H
Assign a unique device address to the oscilloscope. Two devices
cannot share the same device address.
H
Turn on at least two-thirds of the GPIB devices while using the
network.
Summary of Contents for 071-0957-03
Page 4: ......
Page 15: ...Getting Started ...
Page 16: ......
Page 51: ...Application Examples ...
Page 52: ......
Page 80: ...Application Examples 2 28 TDS3000B Series User Manual ...
Page 81: ...Reference ...
Page 82: ......
Page 176: ...e Scope 3 94 TDS3000B Series User Manual ...
Page 177: ...Appendices ...
Page 178: ......
Page 216: ...Appendix E Performance Verification E 2 TDS3000B Series User Manual ...
Page 218: ...Appendix F General Care and Cleaning F 2 TDS3000B Series User Manual ...
Page 234: ...Appendix G Ethernet Setup G 16 TDS3000B Series User Manual ...
Page 235: ...Glossary ...
Page 236: ......
Page 247: ...Index ...
Page 248: ......