Getting Started
TDS3000B Series User Manual
1- 3
Probe Compensation
Perform this adjustment to match your probe to the input channel.
This should be done whenever you attach a passive probe for the first
time to any input channel.
1.
Connect the oscilloscope probe
to channel 1. Attach the probe
tip and reference lead to the
PROBE COMP
connectors,
then push
AUTOSET
.
If using the probe hook-tip,
ensure a proper connection by
firmly twisting the tip onto the
probe.
CH 1
PROBE COMP
AUTOSET
button
2.
Check the shape of the displayed
waveform.
Overcompensated
Compensated correctly
Undercompensated
3.
If necessary, adjust your probe.
Repeat as necessary.
L
P3010
P6139A
NOTE
. See
Appendix D: Probe Basics
for more information about the
probes provided with your oscilloscope.