Appendix D: Probe Basics
TDS3000B Series User Manual
D- 11
Supported Active Probes and Adapters
The oscilloscope supplies power to active probes. You can use the
following active probes as long as the total power required by the
probes does not exceed the capacity of the oscilloscope. To
determine the total probe load, add the load factors for all probes you
want to use. The oscilloscope can supply power to this combination
if the sum of the load factors equals 10 or less. All passive probes
have a load factor of zero.
Active probe
Recommended usage
Load factor
P6205
FET probe, 750 MHz, 10X
0
P6243
SMT probe, 1 GHz, 10X
0
P5205
High-voltage differential probe,
1300 V, 100 MHz, 50X or 500X
6
P5210
High-voltage differential probe,
5600 V, 50 MHz, 100X or 1000X
6
ADA400A
Differential preamplifier, 10
V sensitiv-
ity, DC to 10 kHz
5
AFTDS
Telecom differential 50
Ω
adapter
0
AMT75
Telecom 75
Ω
adapter
0
TCP202
Current probe, 15 A, DC to 50 MHz
4
013-0278-00
Video display clamp
5
CAUTION.
To avoid a measurement error, do not connect active
probes with a combined load factor that is greater than 10. The
signal distortion caused by such an overload can be subtle (reduced
gain, dynamic range, or slew rate).
Summary of Contents for 071-0957-03
Page 4: ......
Page 15: ...Getting Started ...
Page 16: ......
Page 51: ...Application Examples ...
Page 52: ......
Page 80: ...Application Examples 2 28 TDS3000B Series User Manual ...
Page 81: ...Reference ...
Page 82: ......
Page 176: ...e Scope 3 94 TDS3000B Series User Manual ...
Page 177: ...Appendices ...
Page 178: ......
Page 216: ...Appendix E Performance Verification E 2 TDS3000B Series User Manual ...
Page 218: ...Appendix F General Care and Cleaning F 2 TDS3000B Series User Manual ...
Page 234: ...Appendix G Ethernet Setup G 16 TDS3000B Series User Manual ...
Page 235: ...Glossary ...
Page 236: ......
Page 247: ...Index ...
Page 248: ......