1735 HD High Definition Waveform Monitor
5–1
Performance Verification
This sections consists of two separate procedures. First is the Performance
Check, used to determine compliance with the Performance Requirements in the
Specification. Next is the Adjustment Procedure, used to return the instrument to
operation within specifications.
In both procedures, controls and connectors on the 1735 HD front panel and rear
panel are fully capitalized (such as TWOLINE). Control and connector names on
test equipment, as well as internal controls and adjustments for the instrument
under test, are initial capitalized (such as Time/Div).
Recommended Equipment List
The following equipment and accessory items are required to do the Performance
Check and/or Adjustment Procedures. Broad specifications are followed by an
example of equipment that meets these specifications. In most cases, the
following procedures were prepared using the recommended equipment.
Table 5–1: Recommended equipment list
Description
Requirements
Example
Electrical Instruments
Test Oscilloscope
Tektronix 2465B Oscilloscope with an Opt. 05
T Tr
er
d 1
r be
13 .
Vertical Amplifier:
100 MHz Bandwidth, 5 mV Sensitivity.
(TV Trigger), and a 10X probe (P6137).
Time Base:
10 ns/div to 5 ms/div sweep speeds,
triggering to 5 MHz.
Television Signal Generator
The generator must be capable of delivering a
Color Bar test signal, a Five-Step Linearity
signal, a Field Rate Square Wave signal, a
10% to 90% DC Bounce signal, a swept wave
of 0–30 MHz
±
2%, and a Sync signal for the
television standard of the monitor to be
tested.
Tektronix TSG 1001.
Sine-Wave Generator
Frequency Range: at least 250 kHz to
30 MHz.
Tektronix SG503 Leveled Sine-Wave
Generator installed in a Tektronix TM500
Series Power Module.
Function Generator
Signal: capable of a +5 V / –5 V square wave
at 2 kHz.
Tektronix FG 502 Function Generator
installed in a Tektronix TM500 Series Power
Module.
Calibration Generator
Signal: 0–1 V fast rise pulse with trigger
output.
Tektronix FG 506A Calibration Generator
installed in a Tektronix TM500 Series Power
Module.
Summary of Contents for 070-8462-03
Page 5: ......
Page 13: ...Table of Contents viii 1735 HD High Definition Waveform Monitor...
Page 19: ...Preface xiv 1735 HD High Definition Waveform Monitor...
Page 20: ...Getting Started...
Page 21: ......
Page 35: ...Installation 1 14 1735 HD High Definition Waveform Monitor...
Page 36: ...Operating Basics...
Page 37: ......
Page 63: ...Operating Basics 2 26 1735 HD High Definition Waveform Monitor...
Page 64: ...Specifications...
Page 65: ......
Page 75: ...Specifications 3 10 1735 HD High Definition Waveform Monitor...
Page 77: ......
Page 78: ...Theory of Operation...
Page 79: ......
Page 109: ...Theory of Operation 4 30 1735 HD High Definition Waveform Monitor...
Page 110: ...Performance Verification...
Page 111: ......
Page 134: ...Adjustment Procedure...
Page 135: ......
Page 158: ...Maintenance...
Page 159: ......
Page 163: ...Preventive Maintenance 7 4 1735 HD High Definition Waveform Monitor...
Page 179: ...Corrective Maintenance 7 20 1735 HD High Definition Waveform Monitor...
Page 181: ...Repackaging 7 22 1735 HD High Definition Waveform Monitor Figure 7 10 Repackaging...
Page 182: ...Replaceable Electrical Parts...
Page 183: ......
Page 252: ...Diagrams...
Page 253: ......
Page 256: ...Replaceable Mechanical Parts...
Page 257: ......
Page 265: ...Replaceable Mechanical Parts 10 8 1735 HD High Definition Waveform Monitor...
Page 266: ...Appendices...
Page 267: ......
Page 270: ......
Page 271: ......