TM 514 – COMMON RAIL TEST
Error
Code
175
Functional error!
Impossible to finish current process.
Probable Cause
Solution
- Injector flow deficiency.
- Mounting error of the hydraulic connections.
- Leaking hydraulic connections.
- Dirt on the valves causing retention
problems.
- Dirt in the measuring system.
- Steel filings on level sensors.
- Level sensor faults.
- Injector activation driver failures.
- Faults in the power supplies of the drivers.
- Failures in pressure control.
- Injector is not able to perform the entire flow
test.
- Replace the injector and try to retake the test.
- Check all hydraulic connections, mainly the
returning ones. Return connections must be
connected to their respective return channels.
- Check connection leakages.
- Contact Technical Assistance to perform the
complete periodic maintenance.
Error
Code
176
Functional error!
Impossible to test the DRV valve.
Probable Cause
Solution
- Bad contact with the electrical connection of
the pressure-regulating valve to be tested.
- Valve to be tested is damaged or outside
working range. Valves with electrical damage
will not be mechanically tested.
- Damage to the test valve activation driver.
- Damage to the valve detection circuit.
- Check the electrical connections of the
pressure regulator to be tested.
- Check the condition of the valve to be tested
with an ohmmeter.
- Check the correct application of the valve
- Replace the valve to test the test channel.
Error
Code
177
Impossible to perform calibration optical module CH1
Probable Cause
Solution
- Optical module disconnected.
- Burnt transistors.
- LED module misaligned with Optical Module.
- LED module without power.
- Fluid inside the measuring vessel.
- Contact the manufacturer.
Error
Code
178
Impossible to perform calibration optical module CH2
Probable Cause
Solution
- Optical module disconnected.
- Burnt transistors.
- LED module misaligned with Optical Module.
- LED module without power.
- Fluid inside the measuring vessel.
- Contact the manufacturer.