TM 514 – COMMON RAIL TEST
Error
Code
166
Parameterization error - center of mass CH1
Probable Cause
Solution
- CPRC parameterizes SOM board with center
of mass greater than 105mm.
- Contact the manufacturer.
Error
Code
167
Parameterization error - center of mass CH2
Probable Cause
Solution
- CPRC parameterizes SOM board with center
of mass greater than 105mm.
- Contact the manufacturer.
Error
Code
168
Parameterization error - Simultaneous measurement
Probable Cause
Solution
- CPRC parameters SOM board to measure
injection and return simultaneously.
- Contact the manufacturer.
Error
Code
171
Old filter.
Replace and reset hour counter.
Probable Cause
Solution
- Over the limit of hour usage.
- Replace fluid filter by following Tecnomotor
specifications.
- Contact Technical Assistance to perform the
complete periodic maintenance.
Error
Code
172
Old oil.
Replace and reset hour counter.
Probable Cause
Solution
- Test fluid usage hour counter is at the limit.
- Replace test fluid by following Tecnomotor
specifications - Calibration fluid ISO4113.
- Contact Technical Assistance to perform the
complete periodic maintenance.
Error
Code
173
Injector(s) not found.
Connect an injector.
Probable Cause
Solution
- No injector connected to the system.
- Bad contact with injector(s) terminals.
- Injector(s) with short circuit, open or outside
working range. Injectors with electrical damage
will not be mechanically tested.
- Damage to the control board drivers.
- Connect another injector of the same actuator
type (inductive or piezoelectric) to test the
channel.
- Check connections and the condition of the
injector(s).
- Check the condition of the power adapter for
discontinuity.
Error
Code
174
FLASH recording error.
Possibility of losing data.
Probable Cause
Solution
- Error recording memory.
- Call technical assistance.