73
16. Specifications
Bridge voltage (BV)
10 V DC, 2.5 V ±10 % (Current: 60mA max. Remote sensing available)
Singal input range
±3.2 mV/V
Equivalent input/
TEDS
Calibration range
0.3 mV/V to 3.2 mV/V
Calibration
accuracy
Within 0.1 % FS (at 10V for BV and 0.5mV/V or more for sensor sensitivity)
Zero adjustment range
±2.0 mV/V
Accuracy
Nonlinearity
Within 0.01% FS + and 1 digit (at an input of 1m V/V or more)
Zero drift
0.5 μV/˚C or less (RTI)
Gain drift
Within ±0.005 % FS/˚C
A/D conversion rate
100 times/sec
Analog filter
Selectable from 3, 10, 30, 100, 300 Hz (-12 db/oct)
Analog voltage
output
Voltage
Approx. 2 V for 1 mV/V
Load register
2 kΩ or more
TEDS function
IEEE1451.4 Class 2 Mixed-mode interface
Note that the TEDS function cannot be used when remote sensing of excitation
power supply is used.
Display
Characters
14.6 mm high alphanumeric, shown by 7-segment green LED display
Range
-19999 to 99999
Decimal point
Position to be selectable
Refresh rate
4, 6, 10, 20 times/sec
Displayable items
Calibration setting
Zero CAL/Span CAL (TEDS CAL, Actual load CAL, Equiv. input CAL)
Function settings
HI value, LO value, HH value, LL value, Comparison mode, Hysteresis, Near-zero,
Digital filter, Analog filter, Motion detection, Zero tracking, Static strain, Digital zero,
Zero offset, Hold mode, Key lock, Minimum scale, Refresh rate, Bridge voltage, B.C.D
data refresh rate, RS232C, D/A converter
Hold functions
Sample hold, Peak hold, Bottom hold, Peak-to-Peak hold, block-specified hold
(Peak, Bottom, Peak-to-Peak)
External input/
output signal
Input:
Hold, Decision, Clear, Digital zero
Output:
1a contact for each relay: HH, HI, OK, LO, LL, analog voltage
Power supply
AC power rating:
100 to 240 V ±10% 50 - 60Hz AC, 13 W
DC power rating:
10 to 28 V DC, 8 W
Operating temperature range
-10˚C to 40˚C
Storing temperature range
-20˚C to 60˚C
Operating humidity range
85 % RH or less (no condensation)
Standard
Safety standard
Installation class I, EN61010-1 Pollution degree 2 Overvoltage category II,
CE marking
EMC
VCCI (Class A), EN61326 (Class A), EN61000-3-2, EN61000-3-3
External dimensions (W x H x D)
Approx. 96 mm x 96 mm x 146 mm (protrusions not included)
Weight
Approx. 950 g
Option (Not used for option-installed models)
NDIS panel (PN-260ND)
Option-installed models
TD-260T AC(BCD) / TD-260T DC(BCD)
TD-260T AC(232C) / TD-260T DC(232C)
TD-260T AC(D/A) / TD-260T DC(D/A)