62
11-8. Set commands
HI Value (W01)
W 0 1 0 0 0 0 0 0 CR LF
sign bit
Plus: 0
Minus: -
Set value without
decimal point
(5 digits max)
LO Value (W02)
W 0 2 0 0 0 0 0 0 CR LF
sign bit
Plus: 0
Minus: -
Set value without
decimal point
(5 digits max)
Compairson Mode (W03)
W 0 3 0 0 0 0 0 0 CR LF
0: Always compare to generate
decision outputs
1: Compare to generate decision
outputs when stabilized
2: Always compare to generate
decision outputs except near zero
3: Compare to generate decision
outputs when stabilized except
near zero
4: Compare to generate decision
outputs when in hold mode
5: Do not compare
Hysteresis (W04)
W 0 4 0 0 0 0 0 0 CR LF
Set value without
decimal point
(5 digits max)
Digital Offset (W05)
W 0 5 0 0 0 0 0 0 CR LF
sign bit
Plus: 0
Minus: -
Set value without
decimal point
(5 digits max)
Near Zero (W06)
W 0 6 0 0 0 0 0 0 CR LF
Set value without
decimal point
(4 digits max)
HH Value (W07)
W 0 7 0 0 0 0 0 0 CR LF
sign bit
Plus: 0
Minus: -
Set value without
decimal point
(5 digits max)
LL Value (W08)
W 0 8 0 0 0 0 0 0 CR LF
sign bit
Plus: 0
Minus: -
Set value without
decimal point
(5 digits max)
Enable HH and LL (W09)
W 0 9 0 0 0 0 0 0 CR LF
0: Disabel
1: Enable
Digital Filter (W11)
W 1 1 0 0 0 0 0 0 CR LF
0: OFF
1: 4 times
2: 8 times
3: 16 times
4: 32 times
5: 64 times
6: 128 times
7: 256 times
Analog Filter (W12)
W 1 2 0 0 0 0 0 0 CR LF
0: 3Hz
1: 10Hz
2: 30Hz
3: 100Hz
4: 300Hz
Motion Detection Time (W13)
W 1 3 0 0 0 0 0 0 CR LF
Time
Motion Detection Bandwidth (W14)
W 1 4 0 0 0 0 0 0 CR LF
Set value without decimal point
(3 digits max)
Zero Tracking Time Interval (W15)
W 1 5 0 0 0 0 0 0 CR LF
Time
11. RS-232C Interface (TD-260T AC(232C) / TD-260T DC(232C))