17
4-2-2. F2 setup screens
x: Applicable
Setup
screen
number
Setup item name
Initial
setting
Setting range
Page
SET
Lock
CAL
Lock
Unit
Displayed text
F2-1
Hold Mode
1.SPL
0.OFF: Disable hold
43
x
1.SPL: Sample hold
2.PEk: Peak hold
3.bot: Bottom hold
4.P2P: Peak-to-Peak hold
F2-2
Digital Offset
00000
-19999 to 19999
52
x
F2-3
Digital Filter
OFF
Moving average: OFF, 4, 8, 16, 32, 64, 128,
256
39
x
# of
times
F2-4
Analog Filter
100
3, 10, 30, 100, 300
39
x
Hz
F2-5
Bridge Voltage
2.5
10, 2.5
24
x
V
F2-6
Remote Sense/TEDS
0.Aut
0.Aut: TEDS auto-calibration
23
x
1.tEd: Enable TEDS/disable remote sense
2.rSE: Enable remote sense/disable TEDS
F2-7
External Hold Mode
0.LEv
0.LEv: Level, 1.PLS: Pulse
43
x
F2-8
Enable CLEAR Signal
ON
OFF, ON
44
x
4. Setup Mode