26
4. To set the rated capacity;
change the blinking digit of the value by pressing
5
or
b
.
To move to the previous or next digit, press
g
or
t
.
To change the digit, press:
To move to the previous
or next digit, press:
To change the decimal point position, blink the digit
you want to the left of the decimal point by pressing
g
or
t
, and then press FNC.
NOTE
The position of the decimal point you set in this step is used
for readings on the TD-260T.
5. To save the rated capacity you entered, press ENTER.
CAUTION
If you use the D/A option, the rated capacity is set as the D/
A full scale value.
6. To perform zero-point calibration;
with no load on the sensor, press ZERO, and then
press ENTER.
The display shows “0-CAL,” followed by zero for a reading,
showing that the zero-point calibration is completed.
If a calibration error is displayed, make the appropriate
correction, and then repeat the calibration.
6. Calibration