67
12-3. D/A Zero
You can set a value so that the TD-260T generates a D/A zero
output (voltage 0 V and current 4 mA) when it displays the
same reading as the value you set.
1. Enter the setup screen number F4-4.
Press 4 times
Press 3 times
2. To change the blinking digit of the value, press
5
or
b
.
To move to the previous or next digit, press
g
or
t
.
To change the digit, press:
To move to the previous
or next digit, press:
3. To save the setting, press ENTER, and then to exit the
setup mode, press ESC.
To save, press:
12-4. D/A Full Scale
You can set a value so that the TD-260T generates a D/A full
scale output (voltage 10 V and current 20 mA) when it displays
the same reading as the value you set.
1. Enter the setup screen number F4-5.
Press 4 times
Press 4 times
2. To change the blinking digit of the value, press
5
or
b
.
To move to the previous or next digit, press
g
or
t
.
To change the digit, press:
To move to the previous
or next digit, press:
3. To save the setting, press ENTER, and then to exit the
setup mode, press ESC.
To save, press:
CAUTION
The rated capacity is set as the D/A full-scale value when
you save the rated capacity during calibration.
12. D/A Converter (TD-260T AC(D/A) / TD-260T DC(D/A) )