7
8
Set keys
FNC:
Pressing this key while a reading is displayed
changes the display to the setup screen.
Each time you press the key while the setup
screen is displayed, the display cycles through
the function numbers.
Pressing the key while the setup screen is
displayed with a digit of the set value blinking
changes the sign setting or the decimal point
position.
ESC/CLEAR: Pressing this key while the setup screen is
displayed cancels the setting, and then displays
a reading again.
Pressing the key while the TD-260T is in a hold
mode clears the hold value.
ENTER:
Pressing this key saves the item or value you
have set. Pressing and holding the key while
a reading is displayed disables or enables all
the keys on the TD-260T. While all the keys are
disabled, the key input is disabled, except the
ENTER key that can enable all the keys again by
being pressed and held.
5b
:
Pressing these keys selects an item you want to
set or changes a set value.
gt
:
Pressing these keys selects a digit you want to
make changes to.
9
ZERO key
The operation of this key varies depending on the setting
for CAL Lock (F3-5).
Pressing this key performs zero-point calibration while the
calibration lock setting is OFF.
Pressing this key while the calibration lock setting is ON
forces a reading on the display to be zero .
(Digital Zero function)