8-7. Peak hold
8-7-1. Operation without block setting
If the HOLD key is pressed or the HOLD signal from the control input/output terminal A7 is set to ON, the TD-260T starts holding
and displaying the maximum reading (peak value). If the HOLD key is pressed again or the HOLD signal from the control input/
output terminal A7 is set to OFF, the TD-260T exits the Peak hold mode, and then starts measuring the input signal again.
If the CLEAR key is pressed or the CLEAR signal from the control input/output terminal A5 is set to ON while the TD-260T is in the
hold mode, the TD-260T resets the peak value.
Reading
Sensor input
Time
A5 CLEAR
Enable
Reading
level response
HOLD key or
pulse response
A7 HOLD
OFF
ON
OFF
ON
OFF
ON
HOLD
A6 JUDGE
OFF
ON
OFF
ON
Decision output
(with JUDGE signal ON)
8-7-2. Operation with block setting
The TD-260T continues to display the reading that has been displayed when a Peak hold operation is terminated.
If the CLEAR key is pressed or the CLEAR signal from the control input/output terminal A5 is set to ON, the TD-260T resets the
display.
Reading
Sensor input
Time
Reading
OFF
ON
OFF
ON
OFF
ON
HOLD
OFF
ON
OFF
ON
HOLD with block setting
A5 CLEAR
level response
HOLD key or
pulse response
A7 HOLD
A6 JUDGE
Decision output
(with JUDGE signal ON)
Enable
If the CLEAR key is pressed or the CLEAR signal from the control input/output terminal A5 is set to ON while the TD-260T is in the
hold mode, the TD-260T resets the peak value.
8. Hold function