14-2. Restore TEDS Data
You can restore the factory calibration setting (“original TEDS
data”) into the TEDS memory after saving data by using the
function explained in the previous section 14-1. “Save TEDS
Data.”
1. Enter the setup screen number F4-7.
Press 4 times
Press 6 times
2. To enter 00015, change the blinking digit of the
value, by pressing
5
or
b
. To move to the previous or
next digit, press
g
or
t
.
To change the digit, press:
To move to the previous
or next digit, press:
NOTE
<
Be sure to enter this value to prevent the original TEDS data
from being overwritten by mistake.
<
If you press ESC while “tEdS” is not displayed, the setting is
interrupted, and then the TD-260T exits the setup mode.
3. Press ENTER. The TD-260T displays “tEdS” and
starts loading the oritinal TEDS data from the TEDS
memory.
4. The TD-260T displays the rated output (mV/V),
showing that the load operation is completed.
Check the values, and then press ENTER.
NOTE
If you press ESC while “tEdS” is not displayed, the setting is
interrupted, and then the TD-260T exits the setup mode.
5. The rated capacity is displayed. Check the value, and
then press ENTER.
NOTE
If you press ESC while “tEdS” is not displayed, the setting is
interrupted, and then the TD-260T exits the setup mode.
6. Press ENTER. The TD-260T displays “tEdS” and
starts saving the data into another area of the TEDS
memory.
7. The TD-260T displays “F4-7,” showing that the save
process is completed, and then waits for a key to be
pressed.
To exit the setup mode, press ESC.