7-7. Zero tracking
Using zero tracking automatically tracks and corrects deviations from zero point, such as drift.
If a drift is within the bandwidth you set, the function automatically moves a drifted zero point to zero at each set time interval.
Zero tracking is enabled only while both CAL Lock (F3-5) and Enable Digital Zero (F3-1) are set to ON.
Zero tracking is disabled while its time interval and bandwidth are respectively set to 0.0 sec and 0.0.
Zero tracking
bandwidth
0
Reading
Time
Zero tracking time interval
zero tracking enabled
zero tracking enabled
7-7-1. Zero Tracking Time Interval
1. Enter the setup screen number F1-11.
Press once
Press 5 times
2. To change the blinking digit of the value, press
5
or
b
.
To move to the previous or next digit, press
g
or
t
.
To change the digit, press:
To move to the previous
or next digit, press:
3. To save the setting, press ENTER, and then to exit the
setup mode, press ESC.
To save, press:
7-7-2. Zero Tracking Bandwidth
You can set a zero tracking bandwidth for the least three
significant digits of a reading of your device.
1. Enter the setup screen number F1-12.
Press once
Press 4 times
2. To change the blinking digit of the value, press
5
or
b
.
To move to the previous or next digit, press
g
or
t
.
To change the digit, press:
To move to the previous
or next digit, press:
3. To save the setting, press ENTER, and then to exit the
setup mode, press ESC.
To save, press: