49
8-9. Peak-to-Peak hold
8-9-1. Operation without block setting
If the HOLD key is pressed or the HOLD signal from the control input/output terminal A7 is set to ON, the TD-260T starts holding
the maximum reading (peak value) and the minimum reading (bottom value) for each sampling. Then, the TD-260T starts holding
and displaying the maximum difference between those two values. If the HOLD key is pressed again or the HOLD signal from the
control input/output terminal A7 is set to OFF, the TD-260T exits the Peak-to-Peak hold mode, and then starts measuring the input
signal again.
If the CLEAR key is pressed or the CLEAR signal from the control input/output terminal A5 is set to ON while the TD-260T is in the
hold mode, the TD-260T resets the peak-to-peak value.
Reading
Sensor Input
Time
Reading
OFF
ON
OFF
ON
OFF
ON
HOLD
OFF
ON
OFF
ON
P: Peak value
B: Bottom value
0
Enable
A5 CLEAR
level response
HOLD key or
pulse response
A7 HOLD
A6 JUDGE
Decision output
(with JUDGE signal ON)
P
B
B
P
P
8-9-2. Operation with block setting
The TD-260T continues to display the reading that has been displayed when a hold operation is terminated.
If the CLEAR key is pressed or the CLEAR signal from the control input/output terminal A5 is set to ON, the TD-260T resets the
display.
Reading
Sensor Input
Time
Reading
OFF
ON
OFF
ON
OFF
ON
HOLD
HOLD with block setting
OFF
ON
OFF
ON
P: Peak value
B: Bottom value
Enable
A5 CLEAR
level response
HOLD key or
pulse response
A7 HOLD
A6 JUDGE
Decision output
(with JUDGE signal ON)
0
P
B
B
P
If the CLEAR key is pressed or the CLEAR signal from the control input/output terminal A5 is set to ON while the TD-260T is in the
hold mode, the TD-260T resets the peak-to-peak value.
8. Hold function