24
6-1-3. Bridge voltage
You need to set the bridge voltage, which is supplied to a
strain gage sensor.
1. Enter the setup screen number F2-5.
Press twice
Press 4 times
2. To enter 2.5 or 10, press
5
or
b
.
Making sure that the output voltage from the sensor is
relatively high, but that it does not exceed the maximum
excitation voltage specified on the sensor test report at the
same time, choose between the two values.
or
3. To save the selection, press ENTER, and then to exit
the setup mode, press ESC.
To save, press:
CAUTION
<
Setting a voltage that exceeds the sensor’s maximum
excitation voltage may cause the sensor to malfunction.
NOTE
<
The initial setting is 2.5 V.
<
For TEDS calibration, if the bridge voltage setting is higher
than the maximum excitation voltage recorded in the TEDS
memory when TEDS data is loaded, the bridge voltage
setting is changed to a setting equal to or lower than the
maximum excitation voltage.
6-1-4. Minimum scale
You can set the minimum increment value that is used to
display a change in the digital reading.
1. Enter the setup screen number F3-3.
Press 3 times
Press twice
2. To select 1, 2, 5, or 10, press
5
or
b
.
or
3. To save the selection, press ENTER, and then to exit
the setup mode, press ESC.
To save, press:
NOTE
The initial setting is 1.
6. Calibration