27
6-3. Actual load calibration
You can perform a calibration by applying an actual load to the
sensor.
Zero-point calibration
Load calibration
1. To perform zero-point calibration;
with no load on the sensor, press ZERO, and then
press ENTER.
The display shows “0-CAL,” followed by zero for a reading,
showing that the zero-point calibration is completed.
If a calibration error is displayed, make the appropriate
correction, and then repeat the calibration.
2. To perform a calibration by using a load;
apply an actual load to the sensor. Press GAIN, and
then press ENTER.
The value for the rated capacity appears with the least
significant digit blinking.
3. To set the rated capacity;
Now you need to enter the value of the load that is
applied to the sensor. To do so, change the blinking
digit of the value by pressing
5
or
b
. To move to the
previous or next digit, press
g
or
t
.
To change the digit, press:
To move to the previous
or next digit, press:
To change the decimal point position, blink the digit
you want to the left of the decimal point by pressing
g
or
t
, and then press FNC.
NOTE
The position of the decimal point you set in this step is used
for readings on the TD-260T.
4. To save the rated capacity you entered, press ENTER.
If a calibration error is displayed, make the appropriate
correction, and then repeat the calibration.
CAUTION
If you use the D/A option, the rated capacity is set as the D/A
full scale value.
6. Calibration