11-10. Control Commands
Hold ON (CE)
C E CR
(Only with a hold mode setting other than 0.OFF enabled)
Hold OFF (CF)
C F CR
(Only with a hold mode setting other than 0.OFF enabled)
Digital Zero (CG)
C G CR
(Only with both Enable Digital Zero and CAL Lock ON)
Reset digital zero (CH)
C H CR
(With Enable Digital Zero ON)
Reset Hold (CI)
C I CR
(Only in hold mode)
JUDGE contact input ON (CJ)
C J CR
(With Enable JUDGE Signal ON)
JUDGE contact input OFF (CK)
C K CR
(With Enable JUDGE Signal ON)
11-11. Format for transmission mode 1
The TD-260T continuously transmits 20-byte data in the
following format.
G S ,
1
,
2
,
3
,
4
, ± 0 0 0 0 0 0 CR LF
sign bit
Plus: 0
Minus: -
Reading
1
Displayed status
O: Overflow (±LOAD,±Full)
S: Stable
M: Unstable
H: Hold
Status notification priority if more than one status
observed
H > O > (S or M)
2
Zero tracking
A: OFF
T: ON
3
Decision output
H: Hi decision output ON
L: LO decision output ON
O: HH decision output ON
U: LL decision output ON
G: OK decision output ON
( HI, LO, HH, and LL decision outputs OFF)
F: All decision outputs OFF
Output notification priority if more than one output
observed
(O or U) > (H or L or G) > F
4
Near zero
N: OFF
Z: ON