83-530-000 Rev. G
57
GOUT
Turns the output to ON or OFF:
“OUT 1/ON” = turn on
“OUT 0/OFF” = turnoff, clears CV and CC bits in the Status Condition (STAT).
OUT ON will respond with “E07’ if the output cannot be turned on because of a
latching fault (OTP< AC, ENA, SO) shutdown.
GSAV
Save present settings. Same settings as power-down last settings listed in Error!
Reference source not found. Except the address and Baud rate are not saved
Saves to the RAM. These settings are erased when the supply power is switched
off and the new ‘last settings’ are saved.
GRCL
Recall last settings. Settings are from last power-down or from last ‘SAV’ or
‘GSAV’ command. Address and Baud rate are not recalled so communication is
not interrupted.
7.9 SINGLE BYTE COMMANDS
7.9.1
General
Single byte commands are commands in which all the necessary data for the supply to act
upon is contained in a single byte. Single byte commands will be executed immediately by the
supply. If the command requires data to be sent to the HOST PC or IEEE Board (see sections
7.9.4 and 7.9.3.1) that response will be transmitted immediately with no delay due to any soft-
ware overhead. With the exception of the Disconnect from communications command, section
7.9.3.1, commands must be sent by the HOST PC or IEEE Board 2 times in sequence for veri-
fication. All have the most significant bit, D7, set to a logic 1. A CR, carriage return, character is
not included in a single byte command. The RST command will not change any setting made by
a single byte command.
All Single Byte commands will be executed in 1 ms or less. This does not include any response
sent to the HOST/IEEE Board, which is dependent upon the response length and the serial
transmission speed (Baud rate).
7.9.2
Global commands without response
7.9.2.1 Disable MD Mode (MD MODE OPTION REQUIRED)
Disable is the default condition upon power up. The Hex value of the command is 0xA0.
Send it two times in sequence. All supplies, both the currently addressed supply and all
non-addressed supplies, will disable MD Mode as a result of this command.
7.9.2.2 Enable MD Mode (MD MODE OPTION REQUIRED)
Send to enable Multi Drop Mode. The Hex value of the command is 0xA1. Sent it two times
in sequence. When this command is sent, the supply will set SRQ retransmission to the dis-
able state; if you wish it to be enabled you must send the enable command. All supplies,
both the currently addressed supply and all non-addressed supplies, will enable MD Mode
as a result of this command.
7.9.2.3 Disable SRQ retransmission (MD MODE OPTION REQUIRED)
Disable is the default condition upon power up. The Hex value of the command is 0xA2.
Sent it two times in sequence. If the supply sends an SRQ it will only sent it 1 time. All sup-
plies, both the currently addressed supply and all non-addressed supplies, will disable SRQ
retransmission as a result of this command. All status registers will retain their data when
this command is sent.
Summary of Contents for GENESYS 10KW
Page 2: ......
Page 3: ......
Page 31: ...83 530 000 Rev G 28 Fig 4 2 Rear panel connections and controls...