PRO CAN PUMP DUAL (PRO-CPD20)
Rev 1.6
2023/07/31
63/95
Copyright
©
2023 by TAEHA CORP. All Rights Reserved
7.13
Cal page(Hardner)
This screen is for calibrating to set the information required for the sensor. (It will be shipped after
setting is completed.)
List
Description
The calibrated current pressure value is displayed.
The calibrated actual pressure value is displayed as the
DEC value.
Input the calibration value.
➔
SET-min : Calibration minimum pressure value
➔
SET-2 : Maximum calibration pressure value
➔
SV.kpa[kPa] : Input pressure value to calibrate
➔
S.Vraw[DEC] : Input the current value by giving the
pressure value to calibrate