List of checking PRO-CPD20 Front Panel button contact point.
➔
This is the name of the front panel of the PRO-CPD20 that
matches the figure on the left.
➔
UP : CHANGE
➔
DN : INSERT
➔
START : RUN/STOP
List of checking the contact point sensor used for PRO-CPD20.
➔
CYL-LOW : On when the cylinder reaches the bottom limit.
➔
CYL-HIGH : On when the cylinder reaches the top limit.
➔
VENT SENS : ON when liquid is detected in vent hose unit.
➔
COVER SENS : On when the cover vacuum value is
reached(Option).
➔
POSITION : Displays the cylinder position value.
When the cylinder-high sensor is on, the position value is 0,
and the position value increases as the cylinder comes down.