12778474-002-2 Revision F – March 2004
TSG-3800 User’s Guide
55
Installing the TSG-3800
Making Electrical Connections
The Wire Wrap headers/DB-9 connectors are designated J1 (PRI), J2 (SEC), J3
(AUX1), and J4 (AUX2).
Figure 2-6
shows the pinouts for these connectors. Pin 1 of
each connector is connected to Ground and is usually connected to the cable shield
(sleeve). Tip 1 (+) and Ring 1 (–) are the signal input pins used by all Input modules,
and are the pins to which the external reference is connected. For 3-Port modules,
Tip 2 (+) and Ring 2 (–) are used for input port 2. Tip 3 (+) and Ring 3 (–) are used
for input port 3. Tip 2 (+) and Ring 2 (–) are used by the Clock Input module for
connecting an external Fault Signal; for example, Cesium PRS.
Figure 2-6.
Signal Input Connections
Selectable input impedances are provided on the Input modules (see
Table 2-4
). A
BNC- to-WW connector, item number 12012421-000-0, is available for connecting
coaxial cable.
Note:
Input Port labeling on the rear panel of the TSG-3800 may
indicate the references used with previous software versions.
Table 2-4. Input Impedance Selections
Input Type
Selections
Default
DS1 Inputs
100
Ω
3.3 k
Ω
(for bridging)
100
Ω
E1 Inputs
75
Ω
120
Ω
3.3 k
Ω
(for bridging)
120
Ω
Clock Inputs
50
Ω
50
Ω
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com