Operating the TSG-3800
Setting Up Input Reference Controls
92
TSG-3800 User’s Guide
12778474-002-2 Revision F – March 2004
For Frequency and MTIE measurements, you can set a threshold and fault mode for
each of three intervals (100, 1000, or 10000 seconds). If you do not define these
limits, they remain at the default values; see
Default Parameter Values
, on page
108. The MTIE and Frequency command syntax is described in
MTIE Command
,
on page 352, and
Frequency Command
, on page 348.
For example, you can assign a REPORT fault mode for a fractional frequency offset
exceeding 1 x 10E
–8
and a FAIL fault mode for an offset exceeding 1 x 10E
–7
. You
can specify an offset threshold for each time scale (T100, T1000, or T10000) for
both limits.
MCD
T100
432
324
T1000
864
648
T10000
1728
1296
Table 3-3. Frequency Pre-defined Settings
Configuration
Time Scale
Limit 1 (Fail)
Limit 2 (Alarm)
PRS
T100
5000
2000
T1000
500
200
T10000
50
20
OCN
(Factory Default Setting)
T100
5000
2000
T1000
750
300
T10000
60
30
MCD
T100
8000
4000
T1000
800
400
T10000
80
40
Note:
IGNORE or ALARM fault modes are the other settings for L1
and L2. For consistency, use L1 for the more extreme limit and fault
mode.
Table 3-2. MTIE Pre-defined Settings (Continued)
Configuration
Time Scale
Limit 1 (Fail)
Limit 2 (Alarm)