SVAN 973 User Manual
88
Field
Compensation
Digital filter that compensates the free-field (
Free
Field
) or diffuse field (
Dif. Field
) effects.
Manual
Windscreen
Digital filter that compensates the effect of the
SA 22 windscreen.
Manual
Range
Position that enables checking the linear operating
range for the sinusoidal signal. The instrument
uses two ranges
– one for the
Level Meter
,
1/1 Octave,
1/3 Octave
and
RT60
functions (32.0
LEQ(A)
– 128.0 PEAK) and another for the
Dosimeter
function (50.0 LEQ(A)
– 141.0 PEAK).
The calibration factor is always added to the range
limits.
Manual
Measurement
Trigger
Screen that enables configuring triggering of the
measurement/integration
process
with
parameters:
Trigger, Source, Level
and
Gradient
.
Manual
Trigger
Position that switches
Off
or on the measurement
trigger by selecting its type:
Level +, Level
–
or
Gr
.
If the instrument is waiting for the trigger condition,
the appropriate trigger icon is flashing on the
display alternatively with the „play” icon.
Manual
Level +
Type of trigger that starts the 1-second
measurement/ integration under the condition:
value of the RMS result (
Source
) integrated during
0,5 ms is greater than the threshold value (
Level
).
In other cases, the instrument continues checking
the trigger condition every 0.5 ms. During one
measurement cycle the instrument performs as
many 1-second integrations as many seconds the
Integration Period
consists
and
stops
the
measurement cycle.
Manual
Level -
Type of trigger that starts the 1-second
measurement/ integration under the condition:
value of the RMS result (
Source
) integrated during
0.5 ms is lower than the threshold value (
Level
).
In other cases, the instrument continues checking
the trigger condition every 0.5 ms. During one
measurement cycle the instrument performs as
many 1-second integrations as many seconds the
Integration Period
consists
and
stops
the
measurement cycle.
Manual
Summary of Contents for SV 973
Page 157: ...SVAN 973 User Manual Appendixes 157 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 158: ...SVAN 973 User Manual Appendixes 158 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 159: ...SVAN 973 User Manual Appendixes 159 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 172: ...SVAN 973 User Manual Appendixes 172 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 173: ...SVAN 973 User Manual Appendixes 173 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 174: ...SVAN 973 User Manual Appendixes 174 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 185: ...SVAN 973 User Manual Appendixes 185 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...