SVAN 973 User Manual
– Appendixes
112
#D,m,?;
- command cannot be executed
9) #D,f,<address>;
function enables the user to delete directory and its contents (files and
subdirectories):
Response:
#D,f;
- command was executed
#D,f,?;
- command cannot be executed
A.7 F
UNCTION
#5
–
STATISTICAL ANALYSIS RESULTS READ
-
OUT
#5 function enables one to read out the statistical analysis results.
#5
function
format is defined as follows:
#5,p;
where:
p
- the number of the profile (1, 2 or 3)
The device responds, sending the current classes of the statistics in the following format:
#5,p;<Status Byte> <LSB of the transmission counter> <MSB of the transmission counter>
<NofClasses><BottomClass><ClassWidth><Counter of the class> (...) <Counter of the class>
Status Byte
gives the information about the current state of the instrument.
D7
D6
D5
D4
D3
D2
D1
D0
where:
D7 = 0 means "overload does not happen",
= 1 means "overload appeared",
D6 = 1 reserved,
D5 = 0 instantaneous current result (RUN State),
= 1 final result (STOP State),
D0 to D4 reserved bits.
Note: There is not any succeeding transmission in the case when the Status Byte is equal to
zero.
The
transmission counter
is a two-byte word denoting the number of the remaining bytes to be
transmitted. Its value is calculated from the formulae:
Transmission counter = 6+n * (4 * the number of the classes in the statistics)
where:
n
the number of the transmitted statistics. For p = 1, 2 or 3 only one statistic is transmitted (n = 1).
NofClasses
is a two-byte word denoting the number of classes in the statistic.
BottomClass
is a two-byte word denoting the lower limit of the first class (*10 dB).
ClassWidth
is a two-byte word denoting the width of the class (*10 dB).
Counter of the class
is a four-byte word containing the number of the measurements belonging to the
current class.
Note: The bytes in the words are sent according to the scheme <LSByte>..<MSByte>.
Summary of Contents for SV 973
Page 157: ...SVAN 973 User Manual Appendixes 157 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 158: ...SVAN 973 User Manual Appendixes 158 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 159: ...SVAN 973 User Manual Appendixes 159 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 172: ...SVAN 973 User Manual Appendixes 172 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 173: ...SVAN 973 User Manual Appendixes 173 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 174: ...SVAN 973 User Manual Appendixes 174 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 185: ...SVAN 973 User Manual Appendixes 185 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...