![Svantek SV 973 User Manual Download Page 107](http://html1.mh-extra.com/html/svantek/sv-973/sv-973_user-manual_1409271107.webp)
SVAN 973 User Manual
– Appendixes
107
In case of <profile> = 1, 2 or 3 the instrument sends results in the format defined as follows:
#2 [,<aver>],<profile>,Xc,(...);
where
c
is the value of the result
X
or question mark (
?
) if result
X
is not available;
If no results are available the instrument will send:
#2,?;
The codes of the results from the
SLM
mode are defined as follows:
v
under-range flag (ccc equals to 0 when the overload did not occur, 2 when the under-range took
place during the last measurement period but did not occur in the last second of the measurement
and 3 when the under-range took place during the last measurement period and it lasted in the last
second of the measurement);
V
overload flag (ccc equals to 0 or 1);
T
time of the measurement (ccc
– value in seconds);
x
start date of the measurement in format
dd/mm/yyyy
(
dd
– day,
mm
– month,
yyyy
- year)
t
start time of the measurement in format
hh/mm/ss
(
hh
– hour,
mm
– minute,
ss
- second)
P
Lpeak
value (ccc
– the value in dB);
M
Lmax
value (ccc
– the value in dB);
N
Lmin
value (ccc
– the value in dB);
S
L
result (ccc
– the value in dB);
R
Leq
result (ccc
– the value in dB).
U
LE
result (ccc
– the value in dB);
B(k)
Lden
result (ccc
– the value in dB; k – flag determining the kind of the result);
I(nn)
LEPd
result (ccc
– the value in dB, nn – the value of Exposure Time in minutes);
Y
Ltm3
result (ccc
– the value in dB);
Z
Ltm5
result (ccc
– the value in dB);
L(nn)
value L of the nn statistics (ccc
– the value in dB).
g
LR15
result (ccc
– the value in dB;);
G
LR60
result (ccc
– the value in dB;).
Note: In the case of Lden, the value k placed in the parenthesis after the code B, denotes the
kind of the currently measured result. The kind of the Lden result depends on the time during
which the measurements were performed (d denotes day, e denotes evening and n denotes
night). The corresponding values of k parameter and the kind of the measured Lden result are
presented below:
k = 1
Ld result,
k = 2
Le result,
k = 3
Lde result,
k = 4
Ln result,
k = 5
Lnd result,
k = 6
Len result,
k = 7
Lden result.
The codes of the results from the
DOSE METER
mode are defined as follows:
v
under-range flag (ccc equals to 0 when the overload did not occur, 2 when the under-range took
place during the last measurement period but did not occur in the last second of the measurement
and 3 when the under-range took place during the last measurement period and it lasted in the last
second of the measurement);
Summary of Contents for SV 973
Page 157: ...SVAN 973 User Manual Appendixes 157 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 158: ...SVAN 973 User Manual Appendixes 158 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 159: ...SVAN 973 User Manual Appendixes 159 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 172: ...SVAN 973 User Manual Appendixes 172 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 173: ...SVAN 973 User Manual Appendixes 173 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 174: ...SVAN 973 User Manual Appendixes 174 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 185: ...SVAN 973 User Manual Appendixes 185 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...