SVAN 973 User Manual
35
Speed of Source value changing
This position appears when the
G
trigger is chosen. The speed of the
Source
value changing (
Gradient
) can be set in the range from
1 dB/ms
to
100 dB/ms
.
4.3 S
ETTING PARAMETERS FOR PROFILES
–
P
ROFILES
Parameters for three profiles can be set in the
Profiles
screen (in case of
Lev. Met.
function) or in the
Profile x
screens (in case of
Dosimeter
function).
Following parameters can be programmed independently
for each profile: weighting filters
for other than peak
results
calculations
(
Filter
), weighting filters for peak
results calculations (
Filter Peak
) and LEQ detectors type
(
Detector
).
<ENT>
Weighting filter
Next weighting filters for both
Filter
and
Filter Peak
positions can be selected:
Z
according to IEC 61672-1:2013 for Class 2,
A
according to IEC 651 and IEC 61672-1:2013 for Class 2,
C
according to IEC 651 and IEC 61672-1:2013 for Class 2,
B
according to IEC 651 for Class 2,
LF
low frequency filter according to China requirements.
LEQ detector selection
Available LEQ detectors (time constants):
Impulse
,
Fast
and
Slow
.
Time constants are applied always to the
Lmax
,
Lmin
,
L
(SPL),
Ltm3
and
Ltm5
results and to the
Leq
,
LE
(SEL),
LEPd
and
Lden
results in case the
Exponential
LEQ detector is selected in the
General Settings
screen (see Appendix D).
4.4 S
ETTING ALARM THRESHOLDS FOR DOSE METER
–
A
LARM
The
Alarm
position is active only in the
Dosimeter
function and is described in detail in the
DOSIMETER
section of this manual.
<ENT>
4.5 C
ONFIGURING DATA LOGGING
–
L
OGGING
Summary Results
(
L
(SPL),
Leq
,
LE
(SEL),
Lden
,
LEPd
,
Ltm3
,
Ltm5
, 10 x
Ln
,
OVL
,
Lpeak
,
Lmax
,
Lmin
,
EX
,
SD
) and spectra are measured and saved in the file with the step defined by the
Integration Period
parameter as many times as defined by the
Repetition Cycles
parameter (
path:
<Menu> / Measurement / General Settings
).
The instrument enables also additional registration of some results with different step defined by the
Logger Step
parameter (
path: <Menu> / Measurement / Logging / Logger Setup
). Therefore, it is
possible to save in parallel two sequences of measured results
– one for Summary Results (SR) and
another for so called Logger Results or Time History results (TH).
When logging is enabled, selected logger results taken from three independent profiles will be saved
simultaneously with time step down to
100ms
. Recording of logger results to a file is stopped after the
Summary of Contents for SV 973
Page 157: ...SVAN 973 User Manual Appendixes 157 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 158: ...SVAN 973 User Manual Appendixes 158 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 159: ...SVAN 973 User Manual Appendixes 159 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 172: ...SVAN 973 User Manual Appendixes 172 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 173: ...SVAN 973 User Manual Appendixes 173 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 174: ...SVAN 973 User Manual Appendixes 174 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 185: ...SVAN 973 User Manual Appendixes 185 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...