SVAN 973 User Manual
– Appendixes
206
Example:
Let us assume that
L35
is equal to 76.8 dB. It means
that during the measurements the noise level 76.8 dB
was exceeded in not more than 35% of the
observation period.
The cumulative density function for the exemplary
data is presented in Figure on the right side. In order
to determine the
Ln
level one has to draw the
horizontal cursor and find out the crossing point
between the cumulative density function and the
cursor. In the instrument the user can determine 10
statistical levels - from
L01
to
L99
(1% step
of observation period).
The display in the instrument presents only first
statistical level N1 (set to: L01 up to L99).
The statistical level
Ln
value, the profile’s number the
statistics are taken from, the RMS detector (
Lin.
, or
Exp.
:
Fast
,
Slow
or
Imp
.), the filter’s name (
A
,
C
or
Z
) and real time are displayed in the top-right side of
the display in one-result view mode.
Exemplary cumulative density
Summary of Contents for SV 973
Page 157: ...SVAN 973 User Manual Appendixes 157 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 158: ...SVAN 973 User Manual Appendixes 158 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 159: ...SVAN 973 User Manual Appendixes 159 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 172: ...SVAN 973 User Manual Appendixes 172 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 173: ...SVAN 973 User Manual Appendixes 173 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 174: ...SVAN 973 User Manual Appendixes 174 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 185: ...SVAN 973 User Manual Appendixes 185 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...