SVAN 973 User Manual
29
3.2.2 History of calibrations
– Calibration History
The
Cal. History
screen displays list of calibration
records which you can view by pressing the
<Enter>
key.
<ENT>
Note: After calibration, the instrument makes two calibration records with the same date
– one
for the Level Meter and one for the Dosimeter.
The calibration record includes information about the
measurement function for which calibration was
performed (
Level Meter
or
Dosimeter
), type of calibration
(
Factory Calibration
,
By Sensitivity
or
By Measurement
),
date of calibration and calibration factor.
▼
Note: Every time you return to the factory settings without keeping the last calibration (see
Chapter
), the new record appears in the list stating the Factory calibration of the instrument.
3.2.3 Erasing calibration records
– Clear History
To erase all calibration records in the history, choose the
position
Clear Hist.
and press
<Enter>
.
The instrument requests confirmation of the selected
operation.
<ENT>
3.2.4 Post measurement calibration
– Post Calibration
Some regulations require to add information about
calibration performed after measurements to the files with
measurement results created before such calibration.
The latest calibration factor is for information purpose
only since it was not considered during the measurement.
The
Post Cal.
screen allows three options: not to save
(
Off
), save in the last created file (
Last File
) or save in
the files which were created after the previous calibration
(
After Cal.
).
<ENT>
3.2.5 Automatic calibration
– Auto Calibration
The
Auto Cal.
position enables the user to perform
automatic calibration when the sound calibrator is
attached.
In this case, the “Calibration by measurement”
screen will appear automatically. If
Auto Cal.
is switched
off, the user should enter this screen through the
Menu
.
<ENT>
Automatic calibration feature was implemented to make calibration as easy as possible and allow the
user to perform calibration of SV 973 with minimum steps.
Summary of Contents for SV 973
Page 157: ...SVAN 973 User Manual Appendixes 157 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 158: ...SVAN 973 User Manual Appendixes 158 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 159: ...SVAN 973 User Manual Appendixes 159 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 172: ...SVAN 973 User Manual Appendixes 172 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 173: ...SVAN 973 User Manual Appendixes 173 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 174: ...SVAN 973 User Manual Appendixes 174 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 185: ...SVAN 973 User Manual Appendixes 185 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...