SVAN 973 User Manual
– Appendixes
182
C
1.605 V
Z
1.605 V
Measuring frequency range
5
Hz ÷ 11.2 kHz with the
Z
filter (-3 dB)
Centre Frequency Ranges for 1/1 Octave
31.5 Hz
÷ 8 kHz
Centre Frequency Ranges for 1/3 Octave
20 Hz
÷ 10 kHz
RMS detector
•
Digital
“True RMS“ with Peak detection
•
Resolution
0.1 dB
•
Range
327.7 dB
•
Crest Factor
unlimited (for signals in 10 kHz band)
Reference conditions
as per IEC 61260-1:2014
•
Reference frequency
1000 Hz
•
Reference level
114 dB
•
Reference temperature
from +20
°C to +26°C
•
Reference relative humidity
from 35% to 65%
Calibration (electrical)
Calibration level
0.11 V
RMS
(@ 114 dB indication)
Basic accuracy
<
0.2
dB (for the temperature T=+23°C
5°C for
sinusoidal signal 114 dB
RMS
in the band 20
Hz ÷ 10 kHz
with the
Z
input filter)
Measurement error
in the full temperature range
<
0.1 dB when the temperature is from -
10°C to +50°C
for the sinusoidal signal 114 dB
RMS
in the band 20
Hz ÷
10 kHz with the
Z
input filter.
Overload detector
The instrument has the built-in overload detectors. Both A/D converter and input amplifier overload conditions
are detected. The overload in the measurement channel (in its analogue part) and the overload of the
analogue / digital converter are both detected.
The “overload” indication appears when the input signal
amplitude is 0.5
dB above the declared “Peak measurement range”.
Warm-up time / Auto-start delay
1 min. (for 0.1 dB accuracy)
Effect of humidity
< 0.5 dB (for 30%<RH<90% at 40
C)
Effect of magnetic field
below electrical noise level (for 80 A/m and 50 Hz)
Antialiasing filter
Summary of Contents for SV 973
Page 157: ...SVAN 973 User Manual Appendixes 157 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 158: ...SVAN 973 User Manual Appendixes 158 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 159: ...SVAN 973 User Manual Appendixes 159 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 172: ...SVAN 973 User Manual Appendixes 172 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 173: ...SVAN 973 User Manual Appendixes 173 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 174: ...SVAN 973 User Manual Appendixes 174 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 185: ...SVAN 973 User Manual Appendixes 185 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...