SVAN 973 User Manual
41
Note: When a signal
recording is waiting for the level trigger the “level”
icon appears
alternatively with the grey „signal” icon.
/
Gradient trigger
The
Gradient +
trigger starts a signal recording for
Rec. Time
under the
condition: the value of the Leq result (
Source
) averaged by 0.5 ms is greater than
the threshold value (
Level
) the speed of this Source result changing (gradient) is
greater than the gradient threshold (
Gradient
). In other cases, recording doesn’t
start, but if it has been already started it can be continued until the
Rec. Time
has
elapsed. The instrument checks the trigger condition also during the recording
and if the condition is met the recording will continue for another
Rec. Time
.
Integration period trigger
When the
Integr. Per
trigger is selected, a signal recording is triggered every time
the measurement starts, and recording will last minimum
Rec. Time
. If the trigger
condition appears during recording (when
Integration Period
is shorter than
Rec. Time
), from this moment, the recording will continue for the next
Rec. Time
and so on.
Manual trigger
When the
Trig.manual
trigger is selected, the signal
recording starts and ends after pressing simultaneously
the
◄
and
►
keys during the measurement. After
pressing these keys, the screen with the corresponding
message appears. The registration always stops after
period defined by the
Rec
. Time
parameter.
Note: When a signal recording is waiting for the gradient trigger manual
trigger or “integration period” trigger, the flashing “trigger” icon
superimposes on the grey „signal” icon.
The
Filter
position enables the user to choose the broadband frequency filter
during a wave recording:
Z
,
A
,
C
,
B
or
LF
.
The
Sampling
parameter defines the sampling frequency of wave recording:
24 kHz
or
12 kHz
.
The
Signal Gain
position enables the user to choose the gain of the recorded
signal:
0 dB
…
40 dB
.
.
The
Source
position indicates the trigger source. Only one measured result can
be used as a triggering source in all modes, namely the instantaneous LEQ from
the first profile (with appropriate filter and detector), which is denoted here as
Leq(1)
.
The threshold (
Level
) can be set in the range from 24 dB to 136 dB. The
Source
value compares with the
Level
value every 0.5 milliseconds.
Checking triggering condition
The
Tr. Period
parameter
defines the time interval for
checking the triggering condition as:
Logger Step
,
0.5ms
,
100ms
and
1s
.
►
Summary of Contents for SV 973
Page 157: ...SVAN 973 User Manual Appendixes 157 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 158: ...SVAN 973 User Manual Appendixes 158 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 159: ...SVAN 973 User Manual Appendixes 159 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 172: ...SVAN 973 User Manual Appendixes 172 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 173: ...SVAN 973 User Manual Appendixes 173 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 174: ...SVAN 973 User Manual Appendixes 174 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 185: ...SVAN 973 User Manual Appendixes 185 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...