SVAN 973 User Manual
34
When a new measurement cycle begins (after pressing the
<Start/Stop>
key or automatically after
previous measurement cycle stop) the instrument checks a trigger condition every 0.5 ms and starts
1-second integration if condition is met.
After 1-second integration, the instrument
repeats trigger condition checking every
0.5 ms
and
starts
next
1-second
integration if condition is met. The
instrument does it as many times as many
seconds are within the Integration Period
and stops the measurement cycle.
Therefore,
the series
of
1-second
measurements may not be continuous,
and the duration of the measurement
cycle
may
be
longer
than
the
Integration Period.
The measurement can be stopped manually at any moment with the
<Start/Stop>
key. Summary Results
are calculated on the base of series of 1-second results measured during each measurement cycle and
saved in a logger file.
The
Level -
trigger starts the 1-second measurement/integration under the
condition: value of the RMS result (
Source
) integrated during 0.5 ms is lower than
the threshold value (
Level
). In other cases, the instrument continues checking the
trigger condition every 0.5 mc.
This is a mirrored trigger to the
Level +
trigger.
Note: When a measurement is waiting for the level trigger, the flashing
“level” icon superimposes on the „wait” icon.
/
Gradient trigger
The
Gr
trigger starts the 1-second measurement/integration under the
condition: value of the RMS result (
Source
) integrated during 0,5 ms is greater
than the threshold value (
Level
) and the gradient of the Source value is greater
than the gradient threshold value (
Gradient
). In other cases, the instrument
continues checking the trigger condition every 0.5 mc.
This trigger has the same logic as the
Level +
trigger, but the trigger condition requires also the gradient
threshold level to be exceeded.
Note: When a measurement is waiting for the gradient trigger, the
flashing “trigger” icon superimposes on the „wait” icon.
Source result
Only one measured result (
Source
) can be used for checking trigger condition in
the
Level Meter
mode, namely the instantaneous LEQ from the first profile (with
appropriate filter and detector), which is denoted here as
Leq(1)
. This position
cannot be changed.
Threshold level
The threshold (
Level
) can be set in the range from
24 dB
to
136 dB
. The
Source
value compares with the
Level
value every 0.5 milliseconds.
Summary of Contents for SV 973
Page 157: ...SVAN 973 User Manual Appendixes 157 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 158: ...SVAN 973 User Manual Appendixes 158 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 159: ...SVAN 973 User Manual Appendixes 159 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 172: ...SVAN 973 User Manual Appendixes 172 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 173: ...SVAN 973 User Manual Appendixes 173 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 174: ...SVAN 973 User Manual Appendixes 174 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 185: ...SVAN 973 User Manual Appendixes 185 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...