APPENDIX A. REMOTE CONTROL
The
USB 2.0
interface
is the Type C serial interface working with 480 MHz clock which enables one to
control remotely the unit. Its speed is relatively high, and it ensures the common usage of USB-C in most
produced nowadays Personal Computers.
The functions, which are developed in order to control data flow in the serial interfaces, ensure:
- Bi-directional data transmission,
- Remote control of the instrument.
The user, in order to programme the serial interface, has to:
1. send "the function code",
2. send an appropriate data file
or
3. receive a data file.
A.1 I
NPUT
/O
UTPUT TRANSMISSION TYPES
Following basic input/output transmission types (called functions) are available:
#1
input/output of the control setting codes,
#2
read out of the measurement results in the
SLM
mode,
#3
read out of the measurement results in the
1/1 OCTAVE
analysis or
1/3 OCTAVE
analysis mode,
#4
read out of the data file from the internal Flash-disc or RAM memory,
#5
read out of the statistical analysis results,
#7
special control functions,
#9
writing the data file into the internal flash-disk.
#D
read/write the data file from the external memory (SD-card),
A.2 F
UNCTION
#1
–
I
NPUT
/O
UTPUT OF THE CONTROL SETTING CODES
#1 function enables the user to send the control setting codes to the instrument and read out a file
containing the current control state. A list of the control setting codes is given in Tab. A.1. The format of #1
function is defined as follows:
#1,Xccc,Xccc,(...),Xccc;
or
#1,Xccc,X?,Xccc,(...),X?,Xccc;
where:
X
- the group code,
ccc
- the code value,
X?
- the request to send the current X code setting.
The instrument outputs in this case a control settings file for all requests X? in the following format:
#1,Xccc,Xccc,(...),Xccc;
In order to read out all current control settings the user should send to the device the following sequence
of characters:
#1;
The instrument outputs in this case a file containing all control settings given in Tab. A1 in the format:
#1,Xccc,Xccc,(...),Xccc;
Example:
The instrument sends the following sequence of characters as an answer for the mentioned
above request:
Summary of Contents for SV 973
Page 157: ...SVAN 973 User Manual Appendixes 157 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 158: ...SVAN 973 User Manual Appendixes 158 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 159: ...SVAN 973 User Manual Appendixes 159 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 172: ...SVAN 973 User Manual Appendixes 172 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 173: ...SVAN 973 User Manual Appendixes 173 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 174: ...SVAN 973 User Manual Appendixes 174 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 185: ...SVAN 973 User Manual Appendixes 185 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...