SVAN 973 User Manual
43
The
Field Compensation
position
enables you to set the
compensation filter for sound measurements in the free-
field (
Free Field
) or diffuse field (
Dif. Field
) conditions or
disable it for laboratory tests. By default, it is set to
Free Field
.
◄
The
Windscreen
position is active when the field
compensation is on and switches on the compensation
when the windscreen is applied. By default, it is
On
◄
Note: In the Simple interface mode, the Comp. Filter position is hidden, but the instrument
will use settings previously defined in the Advanced mode or default settings (Microphone:
On; Field Comp.: Free Field; Windscreen: On)
.
Note:
For laboratory instrument’s tests, the compensation filters should be set in accordance
with the Appendix C
.
4.8 S
ETTING STATISTICAL LEVELS
–
S
TATISTICAL
L
EVELS
In the
Stat. Lev.
screen, you can define ten statistical
levels, named from
N1
to
N10
, to be calculated,
displayed and saved in a file as Summary results (see
Appendix D).
Default statistical levels have following settings:
1
,
10
,
20
,
30
,
40
,
50
,
60
,
70
,
80
and
90
. All values should be within
the integer range [1, 99]. Each value can be set
independently from others.
<ENT>
4.9 P
ROGRAMMING INSTRUMENT
’
S INTERNAL TIMER
–
T
IMER
The
Timer
function is used for programming the automatic measurement start (and turning the
instrument on if it was turned off) at a given time and day of a week.
Note: In the Simple interface mode, the Timer position is hidden, but the instrument will use
settings previously defined in the Advanced mode or default settings (Timer: Off)
.
The
Timer
position allows you to programme the internal
real-time clock to act as a delayed start timer. The
instrument will be switched on automatically at the
programmed time and will perform the measurement with
the same settings used before the instrument was
turned off with one exception (see below Note).
<ENT>
Note: When Timer is On, measurements will be performed from defined Start to Stop times
because the Repetition Cycles parameter will be changed to Inf (path: <Menu> /
Measurement / General Set.). The last integration may be cut.
Measurement start and stop
The
Start (hh:mm)
and
Stop (hh:mm)
positions
determines times of measurement
’s `start and stop. `
►
Summary of Contents for SV 973
Page 157: ...SVAN 973 User Manual Appendixes 157 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 158: ...SVAN 973 User Manual Appendixes 158 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 159: ...SVAN 973 User Manual Appendixes 159 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 172: ...SVAN 973 User Manual Appendixes 172 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 173: ...SVAN 973 User Manual Appendixes 173 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 174: ...SVAN 973 User Manual Appendixes 174 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 185: ...SVAN 973 User Manual Appendixes 185 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...